High precision measurement of intensity peak shifts in tunable cascaded microring intensity sensors.

Optics letters

PubMedID: 27420483

Prasad PR, Selvaraja SK, Varma MM. High precision measurement of intensity peak shifts in tunable cascaded microring intensity sensors. Opt Lett. 2016;41(14):3153-6.
We demonstrate a method to precisely track intensity peak shifts in tunable cascaded double-microring based refractive index sensors. Without modifications, width of the intensity peak of a tunable cascaded microring device limits the precision of peak-shift measurements and thereby the limit of detection of the sensor. We overcome this limitation by using dual harmonic lock-in detection for precisely determining the position of the intensity maximum. Using this modification, we have demonstrated a reduction in the full width at half-maximum (FWHM) of the intensity peak by a factor of over 1300. We show that such a reduction in FWHM of the peak curve can significantly improve the detection limit of a tunable cascaded microring-based sensor.