Improved microcalcification visualization using dual-energy digital mammography.

Acta radiologica (Stockholm, Sweden : 1987)

PubMedID: 23528569

Tsai CJ, Chen RC, Peng HL, Hsu WL, Lee JJ. Improved microcalcification visualization using dual-energy digital mammography. Acta Radiol. 2013;54(6):614-21.
BACKGROUND
Dual-energy digital mammography (DEDM), involving a combination of high-energy (HE) and low-energy (LE) images, has been investigated as offering a potential improvement in microcalcification detection obscured by overlapping tissue structures.

PURPOSE
To explore the possibility to improve detection of microcalcifications using the DEDM technique.

MATERIAL AND METHODS
Three DEDM protocols were performed by adjusting the effective tube current time product (mAs) of LE image at the same (100%), one half (50%), and one-quarter (25%) of that used in HE image acquisition, named DEDM100%, DEDM50%, and DEDM25%, respectively. A single-energy digital mammography (SEDM) method was also used as a control. A total of 525 regions of interest (ROIs) were used to compare the performance of the DEDM to that of SEDM using free-response receiver-operating characteristic (FROC) and areas under the FROC curve (Az).

RESULTS
All DEDM protocols ranked significantly higher than the SEDM method (P < 0.001). The true-positive fraction was 0.90 for an average of 0.017-0.042 false-positive per image using the DEDM100%, 0.017-0.114 using the DEDM50%, 0.021-0.148 using the DEDM25%, and 0.134-0.422 using the SEDM. The estimated Az values were 0.915-0.940, 0.867-0.935, 0.824-0.930, and 0.567-0.673, respectively.

CONCLUSION
The DEDM50% protocol provided a trade-off benefit between accurate microcalcification detectability and radiation dose for any tissue density. Therefore, the DEDM50% has the potential to minimize excess radiation dose without a negative impact on image quality which could improve earlier diagnosis of breast cancer.